Home ] Up ] Contents ] Feedback ] Website News ] Search ] Technical Tips ] Web Resources ] OSHA Issues ] MEPCOM ] Noise Resources ] Presentations Archive ] About ] Top Page 4 ] About ]

                                              Maico MI24 

 Lipin/Dietz Associates, Inc.


Home
Up
Audiometers
Audiometer Calibration
Audiometer Power Protection
Audiometer Calibration Services
Audiometric Test Rooms
Audiometer Paper & Supplies
Audiometer Parts & Supplies
EarShield™ Earphone Covers
Electroacoustic Ears
Municipal Noise
Occupational Health Services
Occupational Health Software
Occupational Health Software
OHM® Software Services
Otoacoustic Emissions
Parts & Accessories
Quest Noise Dosimeters
Quest Sound Level Meters
Sound Level Meters
Spirometers
Spirometer Data Switch
Vision Testers
T2 Vision Testers

Maico MI24 Impedance Audiometer

Impedance Screener

  • Fast, full automatic tympanometric test
  • Ipsi-and contralateral reflex test (MI24 C) with 4 frequencies
  • Reflex test with adjustable levels from 70 to 105 resp. 70 to 110 (MI24 C) dB
  • Fast integrated printer
  • PC interface and compatible with NOAH

 

Top performance and speed
You demand performance and speed from your impedance screener. Maico provides test repeatability and reliability to meet your most stringent requirements. Results are immediately available on the LCD screen and can be printed automatically on our quiet, fast thermal printer.
Comprehensive testing
The MI 24 and MI 26 perform automatic impedance tests, along with multi-frequency, multi-level reflex screening. Both instruments offer graphical reflex display and gradient calculations. Optional contralateral reflex testing is available with either a button phone or TDH 39.
Additional
testing opportunities
Is office space an issue? With the MI 26 you can bring complete air conduction audiometric testing and impedance screening to your office in one compact instrument. Out of office testing is made easier, too, when carrying only one lightweight instrument in our optional soft-side shoulder bag.
Technical Data
 
Tympanometry Mode
Probe Frequency 226 Hz +/- 2% typical
Probe Intensity 85 dB SPL into a 2cc cavity
Pressure Range + 200 to -400 daPa
Volume Range 0 to 5.0 ml
Accuracy +/-10 or 10 daPa
whichever is greater
Rate Of Sweep Variable, 550 daPa/sec. max.
200 daPa/sec. near peak
Test Time Less then 2 seconds typical
Compliance Ranges 0-1.5 ml or 0-3 ml
(automatic selection)
Reflex Mode
IPSI Frequencies 500, 1000, 2000, 4000 Hz +/- 2%
IPSI Intensities 90, 95, 100, 105 dB HL
(100 dB HL max. at 4000 Hz), on/off ratio greater then 70 dB, rise /fall time 10 msec. typical
CONTRA Frequencies 500, 1000, 2000, 4000 Hz +/- 2%
CONTRA Intensities 90, 95, 100, 105, 110 dB HL
(105 dB HL max. at 4000 Hz)
Note: Contralateral reflexes are optional
Display
LCD type graphical display shows graph of tympanogram, peak compliance, peak pressure, canal volume, gradient, reflex, graph, and conclusion
Dimensions 15" W X 4" H X 13" D
38.1 cm W x 10.2 cm H x 33cm D
Weight 10 lb / 4.5 kg
Standard Accessories
Part Number Part
6613 16 ear tip set
5496 Roll of paper
600-4A Probe
   
Optional Accessories
1035-3002 Carrying case

Pricing

Description SKU # Price
MI24 Impedance Screener with IPSI Reflexes 77700-0001 $2957.00

Shipping

For orders totaling Shipping Charge
$100.00 or less $12.00
$100.01-$1000.00 $20.00
$1000.01-$2500.00 $35.00
$2500.01-$4000.00 $45.00
$4000.01-$6000.00 $60.00
$6000.01-$8000.00 Call for pricing

 

 

Home ] Up ]


Copyright © 2012 Lipin/Dietz Associates, Inc.

 

All material contained herein, unless otherwise noted or linked to entities other than this one, remain property of Lipin/Dietz Associates, Inc. and reproduction of them in any form, without exclusive permission, is a violation of applicable copyright laws. Copyright © 2008 Lipin/Dietz Associates, Inc.
All Rights Reserved

Last modified: October 28, 2015